IQ Sensor Net System WTW IQ Sensor Net VARiON Plus Ammonium Nitrate Sensor

Specifications Solutions
Specifications

Specifications

Specifications Ammonium Measurement Nitrate Measurement
Measuring method Electrochemical Electrochemical
Maximum Configuration Common reference electrode, two measuring electrodes, one compensation electrode Common reference electrode, two measuring electrodes, one compensation electrode
Integrable Electrodes - -
Reference Electrode VARiON®Plus Ref VARiON®Plus Ref
Measuring Electrode VARiON®Plus NH4 VARiON®Plus Ref
Compensation Electrode VARiON®Plus K VARiON®Plus Cl
Measuring range/Resolution NH4-N: 1 ... 2,000 mg/l / 1 mg/l; 0.1 ... 100 mg/l / 0,1 mg/l; NH4+: 1 ... 2,580 mg/l / 1 mg/l; 0.1 ... 129.0 mg/l / 0,1 mg/l NO3-N: 1 … 1,000 mg/l / 1 mg/l; 0.1 ... 100 mg/l / 0,1 mg/l; NO3-: 5 … 4500 mg/l / 1 mg/l; 0.5 ... 450.0 mg/l / 0,1 mg/l
Compensation Ranges K+: 0.1 ... 1,000 mg/l / 0.1 mg/l Cl-: 0.1 ... 1,000 mg/l / 0.1 mg/l
Measuring Accuracy in laboratory standard solutions ± 5 % of measured value ± 0.2 mg/l in standard solutions ± 5 % of measured value ± 0.2 mg/l in standard solutions
Calibration Procedures Matrix adjustment against any reference value, 2-point-calibration possible with multiple standard solution Matrix adjustment against any reference value, 2-point-calibration possible with multiple standard solution
Working Life (typically) Reference electrode: 18 months, measuring and compensation electrode: 18 months (in typical application - municipal sewage plants) Reference electrode: 18 months, measuring and compensation electrode: 18 months (in typical application - municipal sewage plants)
Temperature Measurement and Compensation Integrated NTC thermistor; Range 32 °F … 104 °F (0 °C … +40 °C), Accuracy ±0.5 K, Resolution 0.1 K, t95 < 20 s Integrated NTC thermistor; Range 32 °F … 104 °F (0 °C … +40 °C), Accuracy ±0.5 K, Resolution 0.1 K, t95 < 20 s
Pressure Resistance Maximum 0.2 bar (incl. SACIQ sensor connection cable, with installed electrodes) Maximum 0.2 bar (incl. SACIQ sensor connection cable, with installed electrodes)
Ambient Conditions Operating temperature: 32 °F … 104 °F (0 °C … +40 °C), storing temperature: 32 °F … 104 °F (0 °C … +40 °C) Operating temperature: 32 °F … 104 °F (0 °C … +40 °C), storing temperature: 32 °F … 104 °F (0 °C … +40 °C)
Electrical connections 2-wire shield cable with quick fastener to sensor 2-wire shield cable with quick fastener to sensor
Electromagnetic Compatibility EN 61326, Class B, FCC Class A; Intended for indispensable operation EN 61326, Class B, FCC Class A; Intended for indispensable operation
Certifications CE CE
Mechanical Sensor body: V4A stainless steel 1.4571 Protective cup: POM; Temperature sensor: V4A stainless steel 1.4571 Protection rating: IP 68 (0.2 bar, with installed electrodes); Electrode connector: POM Sensor body: V4A stainless steel 1.4571 Protective cup: POM; Temperature sensor: V4A stainless steel 1.4571 Protection rating: IP 68 (0.2 bar, with installed electrodes); Electrode connector: POM
Weight Approx. 1.48 lb (670 g, without electrode, without sensor connection cable) Approx. 1.48 lb (670 g, without electrode, without sensor connection cable)
Warranty VARiON®Plus 700 IQ: 2 years; Electrodes: 1 year for defects of quality VARiON®Plus 700 IQ: 2 years; Electrodes: 1 year for defects of quality
Solutions

Solutions

Recommended Solutions